Back to Lab Process Guides

Cleanroom Environmental & Contamination Control SOP (ISO Class 6 / 7)

Environmental particle counting, differential pressure cascades, ESD mitigation, and solvent wiping protocols per ISO 14644-1 and EPA QAPP guidelines.

Document Metadata & Facility Classification

  • Document ID: SOP-ENV-2021-17 (Rev 2.0)
  • Lead Specialist / Author: Kadeem Manning (Tier II Vacuum Technician)
  • Facility Suite: Cleanroom Suite B-102 (ISO Class 6 Processing / ISO Class 7 Gowning)
  • Compliance: ISO 14644-1 / EPA QAPP / ALCOA+ Data Integrity
  • Official PDF: Download Cleanroom SOP PDF

1. Cleanroom Pressure & Classification Cascade

[Exterior Non-Controlled Hallway] 
              │
              ▼ (+10 Pa Delta)
[Airlock & Gowning Anteroom] (ISO Class 7 / Class 10,000)
              │
              ▼ (+15 Pa Delta)
[Main Vacuum Deposition Bay] (ISO Class 6 / Class 1,000)

2. Contamination Control Standards & Particle Limits

Cleanroom ZoneISO ClassificationMax Allowed Particulates ($\ge 0.5\ \mu\text{m/m}^3$)Temperature RangeRelative Humidity (RH)Pressure Differential
Gowning AirlockISO Class 7 (Class 10,000)$\le 352,000$$20.0^\circ\text{C} \pm 2.0^\circ\text{C}$$45% \pm 5%$$+10\text{ Pa}$ (rel. to hall)
Deposition BayISO Class 6 (Class 1,000)$\le 35,200$$20.5^\circ\text{C} \pm 1.0^\circ\text{C}$$45% \pm 3%$$+25\text{ Pa}$ (rel. to hall)
Metrology HoodISO Class 5 (Class 100)$\le 3,520$$20.5^\circ\text{C} \pm 0.5^\circ\text{C}$$45% \pm 2%$Laminar Flow ($0.45\text{ m/s}$)

3. Strict Material Ingress Protocol

  1. Pre-Wipe Disinfection: Prior to pass-through transfer, wipe all external chemical containers, tool handles, and wafer carriers with $99.9%$ electronic-grade Isopropanol (IPA) using low-particulate polyester cleanroom wipes.
  2. Prohibited Materials: Cardboard, wood-pulp paper, graphite pencils, cosmetics, food/beverage, and non-ESD static-generating plastics are $100%$ prohibited inside Suite B-102.
  3. Logbook Documentation: All personnel entries, tool transfers, and daily particle counts are recorded in synthetic polypropylene cleanroom notebooks per ALCOA+ protocols.